ETHICON INC. SFXSPI MCL+ UNI UD 18IN 2-0 SA CT-1; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID
|
Back to Search Results |
|
| Catalog Number |
SXMP1B413 |
| Medical Device Problem Code |
Positioning Failure (1158)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported an animal underwent an unknown veterinary procedure on an unknown date and barbed suture was used.During the procedure, complaint: 3 separate sutures were used and all 3 had the same issues.The barbs didn¿t hold in the tissue.They opened an alternative box with a different lot and they didn¿t have the same issue.No adverse patient reactions, the sutures were removed and not left in the patient.Alternative sutures were used to close on the 3 occasions.No adverse patient consequences were reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # : (b)(4).H6 component code: g07002 - device not returned.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained via: (b)(4), case 1: no effects recorded for the patient or procedure.Another suture was used.Complaint product is available for return to be evaluated.(b)(4), case 2: no effects recorded for the patient or procedure.Another suture was used.Complaint product is not available for return to be evaluated.(b)(4), case 3: no effects recorded for the patient or procedure.Another suture was used.Complaint product is not available for return to be evaluated.I have the box of sutures with a few unopened remaining which are available for return.These are the same batch/lot as the 3 affected cases above.The practice have another box which is a different lot and expiry which are not experiencing any issues with.Contact for external person who reported: (b)(6).I have requested the complaints shipper to be sent and will get these ready as soon as it arrives.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected information: d4 udi, h6 component code.
|
| |
|
Search Alerts/Recalls
|
|
|