| Catalog Number |
SSF-265 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
05/08/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported ¿capsular contracture baker grade iii to iv", pain, hardening and implant have risen¿.This record is for the left side.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of "capsular contracture" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iii and iv.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported ¿capsular contracture baker grade "3.5 to iv", pain, hardening and implant have risen¿.Later healthcare professional reported "baker iii", "visible distortion" of the implant that is "higher in position" and "noticeable deformation".This record is for the left side.Device was explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: b5, d6b, h6.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: ¿ capsular contracture: unable to observe as it is not related to the manufacturing process.None of the other observations performed during the device analysis (creases and deformation) are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported ¿capsular contracture baker grade "3.5 to iv", pain, hardening and implant have risen¿.Later healthcare professional reported "baker iii", "visible distortion" of the implant that is "higher in position" and "noticeable deformation".This record is for the left side.Device was explanted and replaced.Device was returned.
|
| |
|