| Catalog Number |
AI-07126 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported "balloon would not inflate".Additional information states that the catheter was replaced to continue therapy.No patient injury or consequence reported.The patient's current condition is reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The lot number noted on the returned original packaging pouch is 16f25f0018.Returned for investigate on was a 6fr 110cm wedge catheter with the original packaging pouch.The sample was returned in the ups shipping box and was loosely packed within the original packaging pouch.Upon return, the inflation lumen stopcock was in the open position.The recommended volume capacity of the balloon is 1.0cc.Upon microscopic inspection, the balloon appeared typical; no damage or abnormalities were noted to the balloon.No condensation was noted within the inflation lumen extension line.No contrast media was noted within the injection lumen extension line.No blood was noted within the interior or the exterior surfaces of the returned sample.No visual damage or abnormalities were noted to the returned sample.The sample was likely cleaned prior to the return.The supplied control stroke syringe was not returned with the sample.The inflation lumen was injected with 1.0cc of air using a lab inventory control stroke syringe.The balloon inflated symmetrically.One side of the balloon measured approximately 5mm.The other side measured approximately 5mm.The balloon did meet specifications of radius ratio less than or equal to 1.5.The inflation lumen was injected with 1.0cc of air using a lab inventory control stroke syringe.The balloon inflated successfully without any difficulty and was noted symmetrical.The balloon deflated in less than 3 seconds when the syringe was removed per specification.No pull away was noted after the tug test.The balloon was placed in water and air was injected into the inflation lumen again.No leak was noted.The injection lumen was aspirated and flushed.No blood or debris was noted.A lab inventory 0.025in guidewire was back loaded through the distal tip.No resistance was noted; the guidewire was able to advance through the injection lumen.No blood or debris was noted.A device history record (dhr) review was conducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.The reported complaint that the "balloon would not inflate" is not confirmed.Upon return, no damage or abnormalities were noted to the returned sample.During the investigation, no leak was noted from the returned catheter/balloon.The balloon inflated and deflated as per specifications.The returned device passed visual and functional test specifications.Based on a review of the device history record (dhr), the product met specification upon release.The root cause of the complaint is undetermined.No further action required at this time.
|
| |
|
Event or Problem Description
|
|
It was reported "balloon would not inflate".Additional information states that the catheter was replaced to continue therapy.No patient injury or consequence reported.The patient's current condition is reported as "fine".
|
| |
|
Search Alerts/Recalls
|