• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I OSSEOTITE TAPERED CERTAIN PREVAIL IMPLANT 4/3 X 11.5MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I OSSEOTITE TAPERED CERTAIN PREVAIL IMPLANT 4/3 X 11.5MM; DENTAL IMPLANT Back to Search Results
Catalog Number XIITP4311
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Hypersensitivity/Allergic reaction (1907); Unspecified Infection (1930); Inflammation (1932)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient had swollen submandibular glands and the tissue was irritated around the implants at tooth locations #19 and #27.The implants were removed due to allergic reaction and infection.Symptoms as a result of the event: pain, inflammation.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).B3: date of event unknown / not provided.E1: address, email address and first/last given name unknown / not provided.Product has been received by zimvie and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This report is being submitted to relay additional information and device evaluation.The following sections are being reported: b4: date of this report was updated.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H3: device evaluated by manufacturer was updated.H6: type of investigation codes were added: 4109, 4111 and 3331.H6: investigation findings code was added: 213.H6: investigation conclusions code was added: 67 and 4310.H10: narrative/data was updated.Zimvie received the reported implant for evaluation.Visual evaluation was performed, the implant had signs of use.There was no damage identified or signs of malfunction that would have contributed to the event.Measurements match drawing.Dhr review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number for similar events and 1 other relevant complaint was identified.Based on the investigation and risk management file review, the most likely root cause determined from the investigation was patient factors.A summary investigation has been completed for infection events identifying that a definitive root cause cannot be identified due to the wide range of external (non-design/ non-manufacturing related) factors potentially impacting the sterility of the implant and its environment.Therefore, based on the available information, device malfunction did not occur.No damage to the implant was identified.The reported event was non-verifiable.Infection is a medical condition and is non-verifiable with all the information.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
 
Event or Problem Description
No additional event information received at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSSEOTITE TAPERED CERTAIN PREVAIL IMPLANT 4/3 X 11.5MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key22956413
Report Number0001038806-2025-02092
Device Sequence Number12875088
Product Code DZE
UDI-Device Identifier00844868022121
UDI-Public(01)00844868022121(17)300113(10)2120023934
Combination Product (Y/N)N
PMA/510(K) Number
K130949
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Catalogue NumberXIITP4311
Device Lot Number2120023934
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/15/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date09/03/2025
Supplement Report FDA Received Date10/09/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/14/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-