W. L. GORE & ASSOCIATES, INC. GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE; STENT, SUPERFICIAL FEMORAL ARTERY
|
Back to Search Results |
|
| Catalog Number |
VBJR061502A |
| Medical Device Problem Code |
Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/12/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Cbas® heparin surface incorporates carmeda heparin manufactured from heparin sodium api, which is covalently bound to the device surface and is essentially non-eluting.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Event or Problem Description
|
|
The following information was reported to gore: on august 12, 2025, a gore® viabahn® endoprosthesis (vsx device) was intended for an unknown anatomical location during treatment of an unknown etiology.After the vsx device was advanced across the lesion, the physician attempted to deploy.However, after deployment was initiated the deployment line got hung up at the turn around and started to bow string the stent.The patient did not experience any adverse consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
Section h6 updated to reflect completion of investigation.The manufacturing records were reviewed and documented in the product history task.The device lot met all pre-release specifications.Product investigation report conclusion: the reported complaint could not be independently confirmed because there were no items returned for evaluation.Therefore, an independent assessment of deployment performance could not be conducted.Procedural deployment of the device can be impacted by different factors including but not limited to zipper integrity, delivery system support or stiffness, or presence of dried fluid on the device or within the catheter dual lumen.The cause for the complaint could not be established with the available information.
|
| |
|