| Catalog Number |
42502006402 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Erythema (1840)
|
| Date of Event |
08/30/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).D10: 42532006702 tibia cemented 5 deg stemmed rt size d lot# 66605029.42522100511 articular surface (mc) right 11mm lot# 65206227.42540000032 all poly patella cemented 32mm dia lot# 66700608.H6: suggested code (annex g)- mechanical (g04) - femur.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the device remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported a patient underwent an initial right total knee arthroplasty.Subsequently, post-operatively reported the incision was red and warm to touch about nineteen days post procedure.Issue was treated with an unknown medication.All products remain implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections have been updated: b4; b5; g3; h2; h3; h6; h10; h11.Visual and dimensional evaluations of the product could not be performed as no product was returned nor were pictures provided.The root cause of the reported event was determined to be unrelated to the implanted zimmer biomet device; therefore, a review of the device history records was not performed.Sterilization certification was found to be conforming.Review of the reported event by a health care professional found: it is expected that a wound heals in stages and should be of normal appearance related to the timeframe since the incision was made.A surgical wound should be well approximated without redness, warmth, swelling and/or purulent drainage for the duration of its healing.The expression 'wound concerns' or 'non-healing wound' would imply that the appearance of the wound deviates from what a surgical wound should appear.It may be red, have drainage, additional pain, warmth and swelling as well as healing time may be delayed.This deviation signifies an alteration in the wound healing process which can be complicated by patient comorbidities such as diabetes, obesity, smoking, and other conditions that are known to slow a person¿s ability to heal.Wound complications can be treated conservatively or more invasively with an irrigation and debridement (i&d) which promotes healing at the site and prevents further complications.Investigation results concluded that the root cause of the reported event is attributed to non-device related factors.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information at time of this report.
|
| |
|
Search Alerts/Recalls
|