Catalog Number TGMR404020
Medical Device Problem Code Device Endoleak (4074)
Health Effect - Clinical Code Clinical Endoleak (4843)
Date of Event 08/22/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
The field sales associate (fsa) reported the following to gore: on (b)(6) 2025, the patient underwent an endovascular procedure to treat an aneurysm utilizing the gore® tag® conformable thoracic stent graft with active control system (ctagac).On (b)(6) 2025, a reintervention was performed to address a type 1a endoleak utilizing the gore® tag® conformable thoracic stent graft with active control system.During the procedure, the initial deployment attempt was unsuccessful.The physician proceeded to open the hatch and employed the back-up deployment mechanism, which resulted in successful deployment.An attempt was made to angle the device downward, resulting in slight movement; however, the exact distance of displacement is unknown.Subsequently, an additional ctagac device was implanted and deployed successfully.The type 1a endoleak was resolved.
 
Additional Manufacturer Narrative
Added b5, g3/g4, h1/h2.
 
Event or Problem Description
On (b)(6) 2025, the patient underwent an endovascular procedure to treat an aneurysm utilizing the gore® tag® conformable thoracic stent graft with active control system (ctagac).On (b)(6) 2025, a reintervention was performed to address a type 1a endoleak utilizing the gore® tag® conformable thoracic stent graft with active control system.During deployment of the first (distal) ctagac, the primary deployment line disconnected from the handle immediately after deployment initiation.Prior to opening the access hatch, secondary deployment was initiated and the secondary deployment line was removed, leaving the device fully constrained.Once the access hatch was opened and the primary deployment line was removed, the device expanded to its full diameter.Minor distal bird-beaking was observed, prompting the physician to perform ballooning using a gore® tri-lobe balloon, which successfully improved wall apposition.It was noted that an attempt was made to angle the device downward, resulting in slight movement; however, the exact distance of displacement is unknown.A second (proximal) ctagac was then deployed to enhance the seal at the proximal extent of the treatment zone, with no issues reported during deployment.Notably, angulation was not attempted during deployment of either device.Upon case completion, the fsa observed that the primary deployment line handle of the first ctagac was missing the connector that joins the deployment line to the handle.Ultimately, the type 1a endoleak was resolved and the patient tolerated the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® TAG® CONFORMABLE THORACIC STENT GRAFT WITH ACTIVE CONTROL SYSTEM
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL WOODY SPRINGS B/P
3450 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
monique nwanze
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key22957394
Report Number2017233-2025-06564
Device Sequence Number18207814
Product Code MIH
UDI-Device Identifier00733132646852
UDI-Public(01)00733132646852(17)271201
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P040043
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberTGMR404020
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/22/2025
Supplement Date Received by Manufacturer10/03/2025
Initial Report FDA Received Date09/03/2025
Supplement Report FDA Received Date10/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/02/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age80 YR
Patient SexUnknown
-
-