Model Number PVS25
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code Aortic Valve Stenosis (1717)
Date of Event 01/29/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Note: the manufacturer received the device's model and serial number as additional information.It was noted though that the prosthesis's size reported in the paper does not match the data provided.The manufacturer will submit a follow up report upon receipt of a clarification on this matter.The manufacturing and material records for the device involved in this event, as they pertain to the reported issue, were retrieved and reviewed by the manufacturer¿s quality engineering.The results confirmed that this prosthesis satisfied all required material, visual, and performance standards at the time of manufacture and release.Based on the information available the cause of the event is related to the structural valve deterioration.As reported in the scientific literature, structural dysfunction is the major cause of failure of bioprosthetic heart valves and the principal underlying pathologic process is cuspal calcification.It should be noted that, based on the information retrieved regarding the clinical history, the patient involved in the reported event presented several risk factors for structural valve deterioration (i.E.Dyslipidemia, hypertension, diabetes mellitus type ii).As a further note, since no diagnostic images were available for evaluation, it was not possible to assess if the perceval valve position was adequate to guarantee a normal functioning or if possible abnormal mechanical load and stress distribution on the prosthesis could have contributed to the early degeneration observed in this case.Ultimately, it should be considered that structural valve deterioration is listed as a possible adverse event in the perceval ifu.The event is, therefore, a known inherent risk of the device.
 
Event or Problem Description
Information obtained from a published paper entitled: " emergency transcatheter aortic valve in sutureless valve replacement for managing cardiogenic shock", by alessandro comis, et al.The abstract indicted the following: ¿sutureless aortic bioprosthesis (sbvs) may pose specific challenges to valve-in-valve (viv) implantation, particularly, in the setting of cardiogenic shock which remains challenging and understudied.Paper reports a patient with a sbvs disfunction, presenting with acute decompensated hf.Despite intra-aortic balloon pump (iabp), clinical deterioration prompted an emergency transcatheter viv.Computer tomography imaging confirmed procedural feasibility.A self-expanding evolut pro+ valve was deployment followed by post-dilatation for under expansion.The procedure restored valve function, and the patient was weaned off iabp.At 30-day follow-up, patient was in nyha class i with complete recovery of left ventricular systolic function (ef 53%).Careful ct planning and device selection are critical.Despite the lack of data in this setting, viv may represent a safe and effective option for high-risk patients in cardiogenic shock (cs)¿.Based on the further information received, a patient with acute episodes of heart failure (hf) presented to the emergency department.His medical history included surgical aortic valve replacement with a perceval valve size m in 2019 for severe aortic stenosis.Comorbidities included hypertension, dyslipidemia, type ii diabetes, and a history of smoking.Transthoracic echocardiography (tte) revealed a severely reduced ejection fraction (ef) and svd of the bioprosthesis, leading to severe stenosis.The patient was referred for redo surgical aortic valve replacement after stabilization of acute hf, given patient's age and a surgical risk profile (sts mortality score 5.67%).During hospitalization at the center, despite optimal medical therapy including intravenous diuretics, the patient experienced worsening heart failure (hf) with cardiogenic shock (cs) requiring vasopressor support and intra-aortic balloon pump (iabp) assistance.A thoracic computer tomography (ct) revealed a severe worsening of the pulmonary congestion with diffuse interstitial edema.Due to the clinical deterioration and a significant increase in surgical risk (sts mortality score 26.9%), also related to a drop in ef (20%), an emergency percutaneous viv was indicated.Final angiography showed good expansion of the thv, correct cusp alignment, no paravalvular leaks, and preserved coronary flow.Following the procedure, iabp support was no longer needed, and vasopressors were gradually weaned off.The patient was discharged in nyha class ii 5 days after viv, predischarge tte showed mildly improved ef (27%) and excellent prosthesis function.Based on the further information received, the perceval prosthesis was implanted on (b)(6) 2020.
 
Additional Manufacturer Narrative
Updated sections: g3, g6, h2, h11.The manufacturer received confirmation from the cardiologist who authored the publication that the valve size reported in the paper was incorrect; the actual valve that the patient had implanted was the perceval s icv1210 (b)(6) size l referenced in the initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCEVAL SUTURELESS AORTIC HEART VALVE
Common Device Name
TISSUE HEART VALVE
Manufacturer (Section D)
CORCYM S.R.L.
via per crescentino sn
saluggia, vc
Manufacturer (Section G)
CORCYM CANADA CORP.
5005 north fraser way
burnaby, bc
Manufacturer Contact
laura mannino
5005 north fraser way
burnaby, bc 
MDR Report Key22961137
Report Number3004478276-2025-00172
Device Sequence Number9515050
Product Code LWR
UDI-Device Identifier08022057014719
UDI-Public(01)08022057014719(240)ICV1210(17)240409
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P150011
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Literature,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/09/2024
Device Model NumberPVS25
Device Catalogue NumberICV1210
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/08/2025
Supplement Date Received by Manufacturer09/10/2025
Initial Report FDA Received Date09/04/2025
Supplement Report FDA Received Date10/06/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/09/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
Patient SexMale
Patient Weight70 KG
-
-