DEPUY IRELAND - 3015516266 STD CTR SCREW METAGLENE 10MM; SHOULDER IMPLANT - SCREW Back to Search Results
Catalog Number 130960000
Medical Device Problem Codes Difficult to Advance (2920); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/13/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: h3, h6 investigation summary: no device associated with this report was received for examination.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Event or Problem Description
It was reported that there were major issues with the collet not allowing glenosphere screw to thread.The collet extraction tool also wouldn't fit into the hole.Had to use a dental pick to rip it out.Surgeon was frustrated and ended up wasting the collett, std csm baseplate, and 15 central screw and putting in legacy baseplate +10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STD CTR SCREW METAGLENE 10MM
Common Device Name
SHOULDER IMPLANT - SCREW
Manufacturer (Section D)
DEPUY IRELAND - 3015516266
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
CORK MFG & MATERIAL WAREHOUSE
loughbeg ringaskiddy
cork
EI  
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key22961549
Report Number1818910-2025-15023
Device Sequence Number9636440
Product Code PHX
UDI-Device Identifier10603295525196
UDI-Public(01)10603295525196(17)320430(10)3800218
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K203694
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number130960000
Device Lot Number3800218
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/13/2025
Initial Report FDA Received Date09/04/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/23/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK SCREWDRIVER
Patient Age75 YR
Patient SexFemale
-
-