CORIN LTD MOBILIT; ECIMA INSERT FOR 28MM HEAD Back to Search Results
Model Number HDM2854
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 01/16/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Per (b)(4) - initial report.Additional information including operative notes, x-rays and the lot codes, has been requested in order to progress with the investigation of this event.And if received, will be provided in a supplemental report upon completion of the investigation.The appropriate device details have been requested and the relevant device manufacturing records will be identified and reviewed.Please note: this report is filed with the fda due to an adverse event experienced with a device that is similar to those placed on the market in the usa, however, this event occured outside of the usa.The submission of this report does not constitute an admission that the device, reporting entity, entity's representative or distributor caused or contributed to this event.
 
Event or Problem Description
Mobilit and trinity revision of the ecima insert and biolox delta ceramic head after approximately 2 weeks due to infection.Note: this is a regularization within the framework of a clinical study following a change in our event monitoring process of sae from clinical study.
 
Additional Manufacturer Narrative
Per comp (b)(4) - final report.Additional information requested was provided.The appropriate devices details have been provided and the relevant device manufacturing and sterilisation records have been identified and reviewed.The devices were manufactured, packaged and sterilised according to specifications at the time of manufacture.The cleaning method, the sterilization method and sterile barrier system used to package our devices have a long history of safe and effective use at corin for a wide range of orthopedic devices and has been validated in accordance with the relevant standards.Infection is a known complication with any invasive surgery.Based on the above, no further investigation can be conducted.As the surgeon confirmed that the link between the devices and the event is excluded, the root cause is considered as external.Corin consider now this case closed.Please note: this report is filed with the fda due to an adverse event experienced with a device that is similar to those placed on the market in the usa, however, this event occured outside of the usa.The submission of this report does not constitute an admission that the device, reporting entity, entity's representative or distributor caused or contributed to this event.
 
Event or Problem Description
Mobilit and trinity revision of the ecima insert and biolox delta ceramic head after approximately 2 weeks due to infection.Note: this is a regularization within the framework of a clinical study following a change in our event monitoring process of sae from clinical study.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOBILIT
Common Device Name
ECIMA INSERT FOR 28MM HEAD
Manufacturer (Section D)
CORIN LTD
the corinium centre
cirencester, gloucestershire GL7 1 YJ
UK  GL7 1YJ
Manufacturer (Section G)
CORIN LTD
the corinium centre
cirencester, gloucestershire GL7 1 YJ
UK   GL7 1YJ
Manufacturer Contact
franck didier
the corinium centre
cirencester, gloucestershire GL7 1-YJ
UK   GL7 1YJ
MDR Report Key22962247
Report Number9614209-2025-00571
Device Sequence Number13228555
Product Code LZO
UDI-Device Identifier05056139239205
UDI-Public05056139239205
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K191831
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberHDM2854
Device Catalogue NumberNOT APPLICABLE
Device Lot Number573479
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/03/2025
Supplement Date Received by Manufacturer09/03/2025
Initial Report FDA Received Date09/04/2025
Supplement Report FDA Received Date11/05/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/27/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BIOLOX DELTA / CERAMIC HEAD: 104.2805 LOT 579525.; BIOLOX DELTA / CERAMIC HEAD: 104.2805 LOT UNKNOWN.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age74 YR
Patient SexFemale
Patient Weight91 KG
-
-