Model Number A01130250
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/10/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3: product analysis of product: (b)(4), lot:10d0231 visual and optical examination identified it appears that part of the tip of the instrument has been broken off.The metal deformation gives indication that the failure was the result of overload.E1: initial reporter is unknown medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Information was received from user facility via a manufacturer representative regarding a device used for spinal therapy.It was reported that both the instrument torx shaft and the driver were stripped.There was no patient involved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDICREA LEGACY TL
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
MEDICREA INTERNATIONAL SA
5389 rte de strasbourg
rillieux la pape,fr 69140
FR  69140
Manufacturer (Section G)
MEDICREA INTERNATIONAL SA
5389 rte de strasbourg
rillieux la pape,fr 69140
FR   69140
Manufacturer Contact
justin ellis
1800 pyramid place
memphis, TN 38132
7635265677
MDR Report Key22962265
Report Number1000432246-2025-00021
Device Sequence Number9635993
Product Code HXX
UDI-Device Identifier03613720210757
UDI-Public(01)03613720210757(17)991231(10)10D0231
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/04/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberA01130250
Device Catalogue NumberA01130250
Device Lot Number10D0231
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/13/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/14/2025
Initial Report FDA Received Date09/04/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/01/2010
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-