ROCHE DIAGNOSTICS ELECSYS ANTI-TSHR; THYROID AUTOANTIBODY TEST Back to Search Results
Catalog Number 08496633190
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/15/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The cobas pro ssu serial number was not provided.Sample material from the patient was requested for further investigation.The investigation is ongoing.
 
Event or Problem Description
There was an allegation of questionable thyroid results for one patient from the cobas pro ssu.Refer to the attachment to the medwatch for all patient data.This medwatch is for the elecsys anti-tshr assay.Refer to the medwatchs with a1 patient identifiers (b)(6) and (b)(6) for the other involved assays.
 
Additional Manufacturer Narrative
Testing of the provided patient sample revealed the presence of a streptavidin interfering factor, which caused the falsely elevated values.Per product labeling for the assay: in rare cases, interference due to extremely high titers of antibodies to analyte-specific antibodies, streptavidin or ruthenium can occur.These effects are minimized by suitable test design.For diagnostic purposes, the results should always be assessed in conjunction with the patient¿s medical history, clinical examination and other findings.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS ANTI-TSHR
Common Device Name
THYROID AUTOANTIBODY TEST
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
natalie catton
9115 hague road
indianapolis, IN 46250
MDR Report Key22962675
Report Number1823260-2025-02729
Device Sequence Number15554666
Product Code JZO
UDI-Device Identifier07613336160329
UDI-Public07613336160329
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K193313
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number08496633190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/15/2025
Supplement Date Received by Manufacturer02/24/2026
Initial Report FDA Received Date09/04/2025
Supplement Report FDA Received Date03/17/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-