Model Number 8801075
Medical Device Problem Codes Incorrect, Inadequate or Imprecise Result or Readings (1535); Device Sensing Problem (2917)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/20/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3, h6: no products have been returned to medtronic for analysis.Codes b17, c20, and d15 are applicable.A0908: applicable to the low accuracy.A0709: applicable to the poor recognition, and difficulty passing verification.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Medtronic received information regarding a navigation system being used during a sacroiliac and thoracolumbar procedure.It was reported that recognition was poor, making it difficult to pass verification.Additionally, the accuracy was low.It was noted that the inaccuracy was caused by all the instruments.A different passive marker was opened and used.There was no delay, and no reported patient impact.
 
Additional Manufacturer Narrative
H2, b5: event details have been updated.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Additional information was received.It was reported that the inaccuracy was "a few" millimeters.
 
Additional Manufacturer Narrative
H3, h6: the spheres were returned for analysis.When attached to a known good instrument and fully seated, 23 of the returned spheres displayed a good geometry error with normal tracking and verification.Two spheres had an uneven reflective surface, possibly from cleaning, and displayed a high geometry error.Three spheres were physically damaged with tool marks on both sides of the spheres, but still displayed a good geometry error.One sphere had lost the back side shell and was unusable.Codes b01, c07, and d02 are applicable to this analysis.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
200 medtronic drive
lafayette 80026
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
200 medtronic drive
lafayette 80026
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
7635265677
MDR Report Key22962964
Report Number1723170-2025-03132
Device Sequence Number9520052
Product Code HAW
UDI-Device Identifier20805832070030
UDI-Public(01)20805832070030(17)270108(10)2501201
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K033621
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/31/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8801075
Device Catalogue Number8801075
Device Lot Number2501201
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/20/2025
Supplement Date Received by Manufacturer09/04/2025
10/27/2025
Initial Report FDA Received Date09/04/2025
Supplement Report FDA Received Date09/04/2025
10/31/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/24/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-