Model Number EU2110X15B
Medical Device Problem Codes Circuit Failure (1089); Power Problem (3010)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/25/2025
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer was previously notified that a ventilator required maintenance and cleaning.There was no harm or injury reported.The device was not in patient use.The device was evaluated by a third-party service center.During the evaluation a service required alarm indicating a power failure occurred.The device's system board was replaced to address the issue.Additionally, non-actionable errors were observed in device's downloaded event log.The machine flow sensor was found to be contaminated and was replaced to address the issue.The device's air inlet foam filter was replaced per preventive maintenance protocol.The device was tested and passed all final testing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY EVO
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key22963586
Report Number2518422-2025-109627
Device Sequence Number9893893
Product Code CBK
UDI-Device Identifier00606959054981
UDI-Public(01)00606959054981
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K181166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/04/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberEU2110X15B
Device Catalogue NumberEU2110X15B
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/12/2025
Initial Report FDA Received Date09/04/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/08/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-