• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH 10.0MM MEDULLARY REAMER HEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH 10.0MM MEDULLARY REAMER HEAD Back to Search Results
Catalog Number 352.100
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/18/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the size 10mm reamer broke inside the femur while reaming.All the broken fragments were retrieved.There was a 20 minutes of surgical delay due to the reported event.The procedure was successfully completed.Patient bone is reported as hard and patient outcome is reported as well.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H.11.Additional narrative: added: d9.H3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: h4.Corrected: g1.H3, h4, h6: the product was returned to j&j medtech orthopaedics for evaluation.The j&j medtech orthopaedics team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that synream reamer head ø10 was broken off, the fracture surface was evaluated and no material issue was identified.Both fragments were received.In addition, general corrosion was noted inside the head.Potential cause for corrosion can be traced to maintenance and/or sterilization procedures, however, with available information the complete potential cause cannot be determined.According to the document cleaning and sterilization instructions se_023827 rev ap the following precautions should be taken: all devices must be thoroughly cleaned and inspected prior to sterilization.Long, narrow lumens, blind holes, moving and intricate parts require particular attention during cleaning and inspection.During cleaning, only use cleaning agents that are labelled for use on medical devices and in accordance with the manufacturer¿s instructions (e.G.Temperature, contact time, and rinse time).Cleaning agents with a used dilution ph of within 7¿9.5 are recommended.Highly alkaline conditions (ph >11) can damage components/devices, such as aluminum materials.Do not use saline, environmental disinfection (including chlorine solutions) or surgical antiseptics (such as iodine- or chlorhexidine-containing products).Do not use a cleaning aid that can damage the surface of instruments such as steel wool, abrasive cleaners or wire brushes.The observed broken condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.A dimensional inspection for the synream reamer head ø10 was not performed due to post manufacturing damage.A functional test was not performed as it is not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the synream reamer head ø10 would contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.As part of j&j medtech orthopaedics quality process, all devices ar manufactured, inspected, and released to approved specifications.There is no indication that a design or manufacturing issue has caused the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review: part: 352.100, lot: 19239, manufacturing site: werk selzach, supplier: (b)(4), release to warehouse date: 03 jun 2008, expiration date: na.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
10.0MM MEDULLARY REAMER HEAD
Common Device Name
REAMER
Manufacturer (Section D)
SYNTHES GMBH
luzernstrasse 19-21
zuchwil 4528
SZ  4528
Manufacturer (Section G)
SYNTHES SELZACH
bohackerweg 5
selzach CO 2425
SZ   2425
Manufacturer Contact
elba bello
1302 wrights lane east
west chester, PA 19380
9083863534
MDR Report Key22964822
Report Number8030965-2025-08992
Device Sequence Number21097816
Product Code HTO
UDI-Device Identifier07611819243729
UDI-Public(01)07611819243729(10)19239
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number352.100
Device Lot Number19239
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/18/2025
Supplement Date Received by Manufacturer09/10/2025
10/08/2025
Initial Report FDA Received Date09/04/2025
Supplement Report FDA Received Date09/11/2025
10/09/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/03/2008
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age48 YR
Patient SexFemale
-
-