As reported by the field clinical specialist, approximately 5 years and 1 month after implant of a 20 mm sapien 3 ultra valve via transfemoral tavr procedure, a valve in valve procedure is being planned.The peak gradients were 92.03 mmhg and mean gradients were 62 mmhg.The valve area index was 0.49 cm2.The leaflets were noted to be thickened.Per medical records received for the patient, on the follow up visit post tavr the patient has heart failure nyha class 1 symptoms and possible syncope episode of unknown etiology and preserved ejection fraction.The patient was found to have severe prosthetic aortic stenosis.Per the cardiology consult, the "patient needs intervention for the aortic valve." the patient is being evaluated for tavr or savr.
|
|
Per the instructions for use (ifu), structural valve deterioration (wear, fracture, calcification, leaflet tear/tearing from the stent posts, leaflet retraction, suture line disruption of components of a prosthetic valve, thickening, stenosis), and device degeneration are known potential risks associated with the use of bioprosthetic heart valves.The ifu cautions that accelerated deterioration of the valve may occur in patients with an altered calcium metabolism.Long-term durability has not been established for the valve.Regular medical follow-up is advised to evaluate valve performance.Structural valve deterioration (svd) may be manifested as stenosis with thickened leaflets.Svd may be mild and not require any intervention or it may be moderate to severe.It can cause the heart to work harder to eject blood from the ventricle.Depending on the severity it could be an indication for valve replacement or medical intervention.Tissue calcification is a common failure mode.The mechanisms for bioprosthetic heart valve tissue calcification are not yet fully understood.Many factors can contribute to the onset and propagation of calcification including patient-related (e.G., patient age, disease state, immune status, and other co-morbidities), pharmacological, and intrinsic properties of the valve itself.It is widely understood that patients with chronic renal disease and prior history of calcific stenosis of the native valve may be predisposed to bioprosthetic calcification.During the manufacturing process, all sapien thv valves are 100% visually inspected for defects and 100% functionally tested for proper coaptation under physiological backpressure conditions prior to release for distribution.Therefore, it is highly unlikely that a manufacturing defect or device malfunction contributed to the event.In this case, there was no allegation or indication a manufacturing non-conformance contributed to this product problem.As noted, patient has history of hyperlipidemia and hypertension.Patients with these comorbidities may have an increased risk of developing valve degeneration.The exact cause of the reported event could not be determined.However, investigation results suggest that progression of the patient disease process may have contributed to the complaint event.A review of edwards lifesciences risk management documentation was performed for this case.The reported event is an anticipated risk of the transcatheter heart valve procedure, additional assessment of this adverse event is not required at this time.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.As such, neither a product risk assessment, nor corrective or preventative actions are required at this time.
|