| Model Number |
B3400060M |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/11/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that high impedances were observed on various contacts of the extension cable during placement of a right subthalamic nucleus (stn) lead with an externalized extension for a study.Multiple impedance tests were performed on two extension cables and two extensions, confirming persistent high impedance specifically on contact 1.Despite using a new extension cable and placing a new extension, high impedances on contact 1 remained unchanged.The physician was aware and the situation was left as is.No patient complications have been reported as a result of this event.
|
| |
|
Event or Problem Description
|
|
Addition information was received from the manufacturer rep indicating that impedance finding resolved on their own the day following the initial report with no known additional actions taken.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: analysis of the extension (sn: (b)(6)) identified that there was a setscrew missing at the distal end of the extension.No significant anomalies found.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|