Model Number 1054655
Medical Device Problem Code No Display/Image (1183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/15/2024
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging a malfunction of a trilogy 100 ventilator's touchscreen.It was reported that the screen saver would not cancel, even after pressing the switch.There was no harm or injury reported.The trilogy 100 was returned to the manufacturer for evaluation and the customer's complaint was confirmed.Although there was airflow being delivered, the whole screen remained completely black, and the display was not viewable.The technician recommended replacing the device's lcd display to address the issue.The root cause has been determined to be a failure in the lcd.The lcd display was returned to the manufacturer's product investigation laboratory for further review.The product investigation laboratory could not duplicate the failure of the lcd display when installed into the trilogy test bed.The lcd display functioned as designed.The display was scrapped.Additionally, not related to the malfunction, the left down button is overly sensitive as if pressed twice when only pressing one time.The front panel board was replaced to address the issue.The technician performed repair tests, alarm check, battery check, run-in tests, and cleaning then confirmed that the unit operated properly.The software was confirmed as 14.1.03.The product investigation laboratory installed the front panel board into a trilogy test bed and pressed all the key buttons numerous times without failure.The product investigation laboratory tested all the buttons, leds and speakers and determined that the front panel board operates as designed and could not duplicate the allegations.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY 100
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key22968702
Report Number2518422-2025-109621
Device Sequence Number9062142
Product Code CBK
UDI-Device Identifier00606959025387
UDI-Public(01)00606959025387
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K083526
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/04/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number1054655
Device Catalogue Number1054655
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/15/2024
Initial Report FDA Received Date09/04/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/11/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1