It was reported that the patient was being seen by a prosthodontist to have the hybrid denture torqued to place when it was noted that the patient had pain around the implant at tooth site #29 when torqueing the abutment screw.The implant was placed and torqued to 50ncm.It was immediately loaded as part of a hybrid lower denture.The patient was assessed by clinician and it appeared on radiograph that the implant had "unzipped".The hybrid denture was removed and the implant at this site had slight mobility.After giving local anesthetic, the implant was removed by unscrewing it by hand.Although implant threads could be felt all around osteotomy, the implant had failed to integrate.Some epithelium was removed from the socket and is being allowed to heal on it's own.There was no soft tissue around the implant at removal and none in the socket.
|
|
This report is being submitted to relay additional information and device evaluation.The following sections are being reported: b4: date of this report was updated.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H3: device evaluated by manufacturer was updated.H6: type of investigation codes were added: 4109, 4111 and 3331.H6: investigation findings code was added: 213.H6: investigation conclusions code was added: 67.H10: narrative/data was updated.Zimvie received the reported implant for evaluation.Visual evaluation of the as returned device identified the implant with signs of use, apparent minor bone / tissue attached to external threads.The implant was identified as reported however, no apparent malfunction was identified with the implant that would contribute to the reported event.Implant matched prints where measured.Functional testing to recreate the reported event could not be performed due to the nature of the device & event.Dhr review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.Based on the investigation and risk management file review, the most likely root cause determined from the investigation is inadequate treatment planning.Therefore, based on the available information, device malfunction did not occur.The reported event was non-verifiable with all the available information.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
|