D4.Medical device expiration date: unknown.H4.Device manufacture date: unknown.Investigation summary: bd was unable to reproduce the customer¿s experience with the bactec product based on our internal procedures and the intended use of the product.Retention samples were visually inspected, tested for viable contamination by sub-culturing on tsa, chocolate, sabouraud and schaedler agars plates, voltage output and gram stain.All results were satisfactory.Batch history record review did not identify any evidence for which the customer submitted the complaint.A complaint history review was conducted and only the current complaint was found relating to the incident lot number and the ¿as reported¿ defect code.Complaint is unconfirmed based on retention samples and batch history record review results.As per product insert a gram-stained smear from culture medium may contain small numbers of nonviable organisms derived from media constituents, staining reagents, immersion oil, glass slides, and specimens used for inoculation.Due to the nature of biological materials in media products and inherent organism variability, the user should be cognizant of potential variable results in the recovery of certain microorganisms.Molecular tests performed on positive blood cultures will detect both viable and non-viable organisms commonly found in culture media.Therefore, molecular test results should be evaluated in conjunction with gram stain results in accordance with standard-of-care practices as well as manufacturer¿s instructions for use.Bd was not able to perform an investigation to the retention samples since batch number is unknown.A complaint history review cannot be conducted relating to the incident lot number and the ¿as reported¿ defect code since batch number is unknown.The batch history record could not be reviewed as the lot number is unknown nonetheless batch history records are always reviewed prior to product release.As per product insert a gram-stained smear from culture medium may contain small numbers of nonviable organisms derived from media constituents, staining reagents, immersion oil, glass slides, and specimens used for inoculation.Due to the nature of biological materials in media products and inherent organism variability, the user should be cognizant of potential variable results in the recovery of certain microorganisms.Molecular tests performed on positive blood cultures will detect both viable and non-viable organisms commonly found in culture media.Therefore, molecular test results should be evaluated in conjunction with gram stain results in accordance with standard-of-care practices as well as manufacturer¿s instructions for use.This mdr is being submitted as part of a retrospective review of records dating april 2023-2025, under capa 11910483.
|