BIOSENSE WEBSTER INC QDOT MICRO; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
|
Back to Search Results |
|
| Catalog Number |
D139402 |
| Medical Device Problem Codes |
Coagulation in Device or Device Ingredient (1096); Device Contamination with Body Fluid (2317)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/27/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster inc., or its employees that the report constitutes an admission that the product, biosense webster inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's ref.# (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a cardiac ablation procedure with a qdot micro which had excessive charring.During ablation, the doctor noticed that charring was visible above the electrode; between tip electrode and electrode 2 at the plastic ring separating the electrodes.There was no consequences for the patient.Per the physician, the charring was considered to be excessive and the doctor estimates the patient risk to be increased.The system did not present any error messages and the physician did not see any product problems during use.There were no issues related to temperature or flow on the catheter.Heparinized normal saline was used and the patient was anticoagulated with activated clotting time (act) of >300 maintained throughout.The patient has not exhibited any neurological symptoms since the procedure was completed.The correct catheter settings were set on the ngen generator and the ngen pump was switching from ¿low¿ to ¿high¿ flow during ablation.
|
| |
|
Additional Manufacturer Narrative
|
|
On (b)(6)-2025, the bwi product analysis lab received the complaint device for evaluation.The product analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# pc-(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported that a patient underwent a cardiac ablation procedure with a qdot micro which had excessive charring.During ablation, the doctor noticed that charring was visible above the electrode; between tip electrode and electrode 2 at the plastic ring separating the electrodes.There was no consequences for the patient.Device evaluation details: the device was returned to johnson & johnson medtech (j&j medtech) for evaluation.Following j&j medtech procedures, a visual inspection, irrigation and temperature and impedance test of the returned device were performed.Visual inspection revealed no damage or anomalies on the device; no char was observed.The device was connected to the generator and an ablation cycle was performed and the device was found working correctly.No temperature or impedance issues were observed.An irrigation test was performed, and the device was flushing correctly, no obstructed holes were observed.No irrigation issues were observed.A manufacturing record evaluation was performed for the finished device lot number and no internal action was found during the review.The char reported by the customer could not be replicated during the product investigation; other issues or circumstances may have occurred during the use of the device that may have affected its performance.The loss of these residues could be related to the decontamination process; however, this cannot be conclusively determined.The instructions for use (ifu) contain the following information: when rf energy is interrupted for either a temperature or an impedance rise (the set limit is exceeded), the catheter should be removed, and the tip should be inspected for char/coagulum that may be present on the tip.If present, do not continue the procedure with the same catheter and replace the catheter.If no char/coagulum is present, flush the tip to ensure the irrigation holes are not plugged prior to reinsertion of the catheter inside the patient body.As part of johnson & johnson medtech's quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
|
| |
|