| Catalog Number |
ASP185 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Suction Failure (4039)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/30/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
After receiving this complaint, we searched for other complaints and we didn't find any other complaints regarding the lot number 9260984.Checking quality database revealed no anomaly in connection with the described defect.The investigation is not yet complete, a follow up report will be submitted on completion of the investigation.
|
| |
|
Event or Problem Description
|
|
According to the available information the product didn't work properly, couldn't aspirate and was leaking.According to the report, the notifier identified a broken device without details.No harm to the patient was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
After receiving this complaint, we searched for other complaint and we didn't find others complaints regarding the lot number 9260984.Checking quality database revealed no anomaly in connection with the described defect our hungarian site performs its investigation and concludes: we have thoroughly reviewed the production documentation related to the reported product issue.Based on our internal records and quality checks, we did not identify any deviations or irregularities during the manufacturing process.Unfortunately, as we have not received a physical sample of the affected product, we are unable to conduct a more detailed investigation at this time.A rmf evaluation was performed based on criq206, risk identified 11405 (hazardous situation: no irrigation and/or suction).The evaluation has showed that risks are adequately controlled and reduced as far as possible in the state of art level.Based on this, we can conclude that residual risks are adequately controlled and reduced as far as possible, and the residual risks associated with the use of the product are acceptable when weighed against the benefits to the patient/user.It is concluded that the risks identified are still acceptable and considered as safe.
|
| |
|
Event or Problem Description
|
|
According to the available information the product didn't work properly, couldn't aspirate and was leaking.According to the report, the notifier identified a broken device without details.No harm to the patient was reported.
|
| |
|