• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC SOUNDSTAR ECO 8FG ULTRASOUND CATHETER, SOUNDSTAR ECO 8F ULTRASOUND CATHETER; CATHETER, ULTRASOUND, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC SOUNDSTAR ECO 8FG ULTRASOUND CATHETER, SOUNDSTAR ECO 8F ULTRASOUND CATHETER; CATHETER, ULTRASOUND, INTRAVASCULAR Back to Search Results
Catalog Number 10439011
Medical Device Problem Codes Material Separation (1562); Manufacturing, Packaging or Shipping Problem (2975)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/11/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent an idiopathic ventricular tachycardia ablation procedure with a soundstar eco 8fg ultrasound catheter and the "casing on the distal end of the soundstar® catheter" appeared to have "plastic shaving off the side," like it was "ripped off." the soundstar® catheter was replaced and the issue was resolved.The procedure continued.No patient consequences were reported.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc.Or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
 
Additional Manufacturer Narrative
On 9-oct-2025, the product investigation was completed as the complaint device was not returned.An analysis of the product could not be performed since a physical sample was not received for evaluation.A manufacturing record evaluation was performed for the finished device lot number e8647777 and no internal action related to the complaint was found during the review.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Based on the information available, there is no indication that a design or manufacturing issue has caused the reported complaint condition.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOUNDSTAR ECO 8FG ULTRASOUND CATHETER, SOUNDSTAR ECO 8F ULTRASOUND CATHETER
Common Device Name
CATHETER, ULTRASOUND, INTRAVASCULAR
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
SIEMENS LTD. SEOUL
venture bldg 2, office 112 601
pohang technopark, jigok-dong
gyeongbuk 790-8 34
KS   790-834
Manufacturer Contact
henry boden
31 technology dr
irvine, CA 92618
9497882078
MDR Report Key22983327
Report Number2029046-2025-03027
Device Sequence Number21219872
Product Code OBJ
UDI-Device Identifier10846835008852
UDI-Public(01)10846835008852(17)260619(10)E8647777
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K140318
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/19/2026
Device Catalogue Number10439011
Device Lot NumberE8647777
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/11/2025
Supplement Date Received by Manufacturer10/09/2025
Initial Report FDA Received Date09/05/2025
Supplement Report FDA Received Date10/22/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NGEN GENERATOR.
Patient SexUnknown
-
-