| Model Number |
Z-800F IV INFUSION PUMP |
| Medical Device Problem Code |
Pressure Problem (3012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/05/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Intuvie received a report indicating that an infusion pump failed the 30 psi occlusion test, recording a pressure of 18.2 psi which is outside the acceptable range of 22 to 38 psi.A review of the device's serial number history revealed no similar complaints.The failure mode was confirmed, and the probable cause was determined to be an out-of-calibration condition.The issue was successfully resolved by recalibrating the 30 psi calibration value from 217 to 187.Capa-zm2023-23 was initiated to investigate the root cause for this failure mode.Reference to complaint#: (b)(4).
|
| |
|
Event or Problem Description
|
|
Intuvie received a report indicating that an infusion pump failed the 30 psi occlusion test, recording a pressure of 18.2 psi which is outside the acceptable range of 22 to 38 psi.No patient involvement was reported.
|
| |
|
Search Alerts/Recalls
|