ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ELLIPSE VR; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number CD1411-36C
Medical Device Problem Codes Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574); Capturing Problem (2891)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient experienced inappropriate shock therapy.It was noted that the implantable cardioverter defibrillator (icd) and the right ventricular (rv) lead exhibited unspecified oversensing and high capture threshold.The icd and the rv lead were explanted and replaced.The patient's condition was unknown.
 
Event or Problem Description
New information received indicates that initially the patient presented via remote transmission.The implantable cardioverter defibrillator (icd) exhibited noise oversensing.The patient was in stable condition following the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELLIPSE VR
Common Device Name
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key22986417
Report Number2017865-2025-99426
Device Sequence Number15696364
Product Code NVZ
UDI-Device Identifier05414734507707
UDI-Public(01)05414734507707(10)A000010369(17)170630
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
P910023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2017
Device Model NumberCD1411-36C
Device Catalogue NumberCD1411-36C
Device Lot NumberA000010369
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/17/2025
Supplement Date Received by Manufacturer09/08/2025
Initial Report FDA Received Date09/05/2025
Supplement Report FDA Received Date09/15/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/19/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age49 YR
Patient SexMale
-
-