| Model Number |
7300TFX |
| Medical Device Problem Code |
Device Stenosis (4066)
|
| Health Effect - Clinical Codes |
Dyspnea (1816); Fatigue (1849)
|
| Date of Event |
08/12/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Updated sections: g3, g6, h6 type of investigation, investigation findings, and investigation conclusions.The device history record was reviewed and shows that this device met all manufacturing specifications for product release prior to distribution.No issues were identified that would have impacted this event.Stenosis is a common manifestation of bioprosthetic valve and valved conduit dysfunction and has many potential root causes, including structural valve deterioration (svd), non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis.Bioprosthetic device stenosis is typically related to patient and/or procedural factors and is not typically an indication of a device malfunction related to a manufacturing deficiency.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.The most likely root cause is patient factors, including chronic kidney disease, coronary artery disease, and hyperlipidemia.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated sections: b2, b5, h6 health effect - impact code.
|
| |
|
Event or Problem Description
|
|
It was reported and per investigation that a patient with a 31mm 7300tfx mitral valve underwent a valve-in-valve procedure after an implant duration of 6 years, 11days due to stenosis.The patient presented with dyspnea and fatigue.The procedure was performed successfully with a 29mm 9755rsl transcatheter valve.
|
| |
|
Additional Manufacturer Narrative
|
|
The subject device is not available for evaluation, as it remains implanted in the patient.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
| |
|
Event or Problem Description
|
|
It was reported and per investigation that a patient with a 31mm 7300tfx mitral valve has been placed under evaluation/consideration for a valve-in-valve procedure after an implant duration of 5 years, 11 months due to stenosis.The patient presented with dyspnea and fatigue.Patient is scheduled for procedure.
|
| |
|