• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP AG AS VEGA SYSTEM® PS; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG AS VEGA SYSTEM® PS; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number NX055Z
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Serious Injury
Event or Problem Description
The patient underwent revision total knee arthroplasty of the right knee for loosening of prior components.A midline incision and medial parapatellar approach were performed, and the femoral and tibial bone were prepared using standard instruments.The femoral and tibial components were implanted using third-generation cement technique (vacuum sealed).A tibial tray and polyethylene insert were placed, followed by patellar button placement.During the procedure, bone cuts, trailing, and implantation proceeded as expected with no reported intraoperative complications.Final inspection confirmed appropriate fit and stability of the components, with no evidence of malalignment or improper fixation.The patient tolerated the procedure, was awakened from anesthesia, and transferred to pacu in stable condition.
 
Additional Manufacturer Narrative
The adverse event is filed under ais reference no.(b)(4).Investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Additional Manufacturer Narrative
The adverse event is filed under ais reference no.(b)(4).Manufacturing site evaluation: the device was not returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation of the unit was performed and no parts were returned for failure analysis.Therefore, the investigation was not able to confirm a device issue that could be associated with the reported event.The device history records (dhr) were reviewed for the available lot number; the device was found to be within specification at the time of production.All device history records (dhr) are reviewed and released according to documented procedures and a device is not released if it does not meet requirements or is nonconforming.A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
 
Additional Manufacturer Narrative
The adverse event is filed under ais reference no.(b)(4).Additional information: b5 - description updated.B6 - test results.B7 - patient medical history.D6 - explant date.
 
Event or Problem Description
Update: revision surgery and implantation of non-aesculap devices was performed on (b)(6) 2025.Ontracture of the right knee and ankylosis postoperatively was found, and manipulation of the knee joint under general anesthesia was performed on (b)(6) 2025.
 
Additional Manufacturer Narrative
The adverse event is filed under ais reference no.(b)(4).Additional information: b6, b7.Investigation results: the complained medical device was not made available for investigation.The investigation was based upon batch history review, historical data analysis, event description and review of a pictures.The provided pictures show the following situation: the femoral component (nx033z) as well as the tibial component (nx055z) shows only less and partial bone cement on the intended area.The meniscal component is still fixed on the tibial component.The device quality and manufacturing history records have been checked for the available lot number and were found to be according to the manufacturer's specifications, valid at the time of production.Review of the complaint history revealed that there are no similar complaints filed against the affected batch number.Conclusion/preventive measures: on the basis of the current information and without the complained medical device being available for investigation, a clear root cause cannot be drawn.In the event that the complained medical device will be provided for investigation in the future, an update of this report will be provided unsolicited.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AS VEGA SYSTEM® PS
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
MDR Report Key22993252
Report Number3005673311-2025-00114
Device Sequence Number8833612
Product Code JWH
UDI-Device Identifier04046963810450
UDI-Public(01)04046963810450(11)221108(17)321107(10)52799308
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 04/06/2026
Report Date (Section F) 09/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberNX055Z
Device Catalogue NumberNX055Z
Device Lot Number52799308
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date08/14/2025
Device Age29 MO
Event Location Hospital
Date Report to Manufacturer08/14/2025
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Not provided
Not provided
Initial Report FDA Received Date09/08/2025
Supplement Report FDA Received Date10/01/2025
12/09/2025
04/06/2026
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
Patient SexFemale
Date Report Sent to FDA09/08/2025
-
-