• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - ALCON PRECISION DEVICE ALCON CLOSURE SYSTEM (ACS), PAIR PAK, POLYPROPYLENE; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - ALCON PRECISION DEVICE ALCON CLOSURE SYSTEM (ACS), PAIR PAK, POLYPROPYLENE; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE Back to Search Results
Catalog Number 8065304901
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/14/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported that an ophthalmic suture was broke before cataract surgery.The issue was resolved after replacing the product with another one.There was no patient harm.
 
Additional Manufacturer Narrative
Additional information provided in sections d9, h3, h6 and h11 one opened suture, in a pouch, in a blister, in a bag was received for the report of suture broken.Sample was visually inspected and found to be nonconforming, suture was observed broken.A dimensional inspection was done for suture diameter, and sample was found to be conforming.Functional testing for suture strength was performed and found to be conforming.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The investigation conducted a non-conformance review of the reported lot number.No deviations were identified, and all corresponding production release specifications defined in the device master record were met.Based on the evaluation of the information and materials received, the investigation concluded that the root cause of the reported event is not determined as the returned sample was conforming for all dimensional and functional testing.The returned sample was found to be non-conforming visually with a broken suture, however the sample was conforming for all functional and dimensional testing associated with the reported event.Therefore, a suture broken as described in the complaint was not confirmed and a root cause cannot be determined for the complaint as described by the customer.No further investigative or manufacturing actions are warranted at this time due to that the returned sample met specifications functionally and dimensionally.Suture material samples are tested at incoming for tensile strength and diameter and visually inspected for any defects such as discoloration and frays.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all manufacturer products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALCON CLOSURE SYSTEM (ACS), PAIR PAK, POLYPROPYLENE
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE
Manufacturer (Section D)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer (Section G)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key22993966
Report Number2523835-2025-00946
Device Sequence Number8833638
Product Code GAW
UDI-Device Identifier00380653049015
UDI-Public(01)00380653049015(17)280731(11)230831(10)164U8H
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P890002
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065304901
Device Lot Number164U8H
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/15/2025
Supplement Date Received by Manufacturer12/16/2025
Initial Report FDA Received Date09/08/2025
Supplement Report FDA Received Date12/30/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/31/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-