• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. THREE PEG PATELLA 32MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. THREE PEG PATELLA 32MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 200-02-32
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Pain (1994); Osteolysis (2377)
Date of Event 08/22/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent an initial total knee replacement on the right side.Subsequently, the patient experienced radiolucent osteolysis and pain.As a result, approximately four years post-surgery, the patient underwent a revision.Patient was last known to be in stable condition following the event.No further impact to the patient was reported.No additional information is available at this time.This is report 1 of 2 for this event/patient.
 
Additional Manufacturer Narrative
D10: 02-022-51-3513 truliant tib imp crc insert sz 3.5, 13mm: sn# (b)(6), 02-020-13-0335 - truliant cr cem fem cr cem right sz 3.5: sn# (b)(6), 02-022-45-3525 - truliant tib fit tray cem sz 3.5f / 2.5t: sn# (b)(6).The revision reported was likely the result of prosthesis wear of the tibial insert and inclusion of the polyethylene implants in the packaging recall.Additional reasons for the reported revision may be prosthesis wear of the patellar component, but this cannot be confirmed from the provided information.Possible causes for polyethylene damage include malalignment between implants, third body wear, patient-related conditions, inclusion of the polyethylene in the packaging recall, or any combination of these possibilities.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and relevant clinical information were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THREE PEG PATELLA 32MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66 court
gainesville, FL 32653
MDR Report Key22998675
Report Number1038671-2025-02889
Device Sequence Number8997779
Product Code JWH
UDI-Device Identifier10885862039590
UDI-Public10885862039590
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K932690
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/18/2026
Device Catalogue Number200-02-32
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/22/2025
Initial Report FDA Received Date09/08/2025
Date Device Manufactured01/20/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age68 YR
Patient SexFemale
-
-