MAUDE Adverse Event Report: ARTHREX, INC. SUREFIRE SCORPION NEEDLE; MANUAL INSTR, GENERAL SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. SUREFIRE SCORPION NEEDLE; MANUAL INSTR, GENERAL SURGICAL Back to Search Results
Model Number SUREFIRE SCORPION NEEDLE
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/19/2025
Type of Reportable Event Malfunction
Event or Problem Description
On 08/19/2025, it was reported by an arthrex subsidiary employee via (b)(4) that an ar-13991n scorpion needle when passing the needle through the forceps, it does not take sutures and ends up breaking the tip of the needle.This occurred during use in a case with no patient effect.
 
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.Complaint allegation is not confirmed.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.Dfu-0392-sub-eo warns against the usages listed to help avoid needle breakage and potential patient injury.The most likely cause the reported failure can be attributed to user error of the device due to excessive force being used during firing of the needle thus, breaking the needle.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUREFIRE SCORPION NEEDLE
Common Device Name
MANUAL INSTR, GENERAL SURGICAL
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key22999796
Report Number1220246-2025-03757
Device Sequence Number9469546
Product Code MDM
UDI-Device Identifier00888867015920
UDI-Public00888867015920
Combination Product (Y/N)N
Initial Reporter CountryCI
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 10/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSUREFIRE SCORPION NEEDLE
Device Catalogue NumberAR-13991N
Device Lot Number15334933
Was Device Available for Evaluation? No
Distributor Facility Aware Date08/19/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/19/2025
Supplement Date Received by Manufacturer10/23/2025
Initial Report FDA Received Date09/09/2025
Supplement Report FDA Received Date10/29/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-