| Model Number |
420205-16 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
08/21/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An investigation is in progress to determine the cause an rma was issued to evaluate the intuitive surgical, inc.(isi) device.Of this reported event.Additional information is being gathered to determine the contribution of the device to the customer reported issue.
|
| |
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted surgical procedure, the pulley was broken in the articulation.The procedure was completed with no reported injury.
|
| |
|
Event or Problem Description
|
|
Refer to h11 for follow-up information.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation was completed to determine the cause of this reported event.Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The instrument was analyzed and found to have a frayed grip cable.The complaint was confirmed by failure analysis, which indicates that the device may have contributed to the customer reported issue.The probable root cause of a frayed cable is attributed to damage to the cable through external collisions, during use, or during reprocessing.This issue can be resolved by using an alternate instrument to complete the procedure.
|
| |
|