• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC ENDOWRIST;DAVINCI SI; FENESTRATED BIPOLAR FORCEPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC ENDOWRIST;DAVINCI SI; FENESTRATED BIPOLAR FORCEPS Back to Search Results
Model Number 420205-16
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/21/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation is in progress to determine the cause an rma was issued to evaluate the intuitive surgical, inc.(isi) device.Of this reported event.Additional information is being gathered to determine the contribution of the device to the customer reported issue.
 
Event or Problem Description
It was reported that during a da vinci-assisted surgical procedure, the pulley was broken in the articulation.The procedure was completed with no reported injury.
 
Event or Problem Description
Refer to h11 for follow-up information.
 
Additional Manufacturer Narrative
An investigation was completed to determine the cause of this reported event.Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The instrument was analyzed and found to have a frayed grip cable.The complaint was confirmed by failure analysis, which indicates that the device may have contributed to the customer reported issue.The probable root cause of a frayed cable is attributed to damage to the cable through external collisions, during use, or during reprocessing.This issue can be resolved by using an alternate instrument to complete the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOWRIST;DAVINCI SI
Common Device Name
FENESTRATED BIPOLAR FORCEPS
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 95051
4085232100
MDR Report Key22999817
Report Number2955842-2025-37273
Device Sequence Number21185551
Product Code NAY
UDI-Device Identifier00886874111642
UDI-Public(01)00886874111642(11)240613(10)K10240613
Combination Product (Y/N)N
Initial Reporter CountryEC
PMA/510(K) Number
K050369
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign
Initial Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 11/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number420205-16
Device Catalogue Number420205
Device Lot NumberK10240613 690
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/21/2025
Supplement Date Received by Manufacturer11/13/2025
Initial Report FDA Received Date09/09/2025
Supplement Report FDA Received Date11/17/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/13/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DA VINCI INSTRUMENTS AND ACCESSORIES
Patient SexUnknown
-
-