Catalog Number TSVM6B8
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/21/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).A4: weight unknown / not provided.G4: additional pma/510(k) number ¿ k101880.Product has been received by zimvie and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the implant at tooth site #19 was fractured.The implant was removed and the patient was placed on an antibiotic.
 
Event or Problem Description
No additional event information received at the time of this report.
 
Additional Manufacturer Narrative
This report is being submitted to relay additional information and device evaluation.The following sections are being reported: b4: date of this report was updated.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H3: device evaluated by manufacturer was updated.H6: type of investigation code was added: 10, 4109, 3331 and 4111.H6: investigation findings code was added: 3252.H6: investigation conclusions code was added: 4307.H10: narrative/data was updated.Zimvie received one the reported device for evaluation.Visual evaluation of the as returned device identified the implant with signs of use, apparent bone/tissue attached to external threads.The implant was observed fractured at the collar region.Functional testing to recreate the reported event could not be performed due to the nature of the device & event.This complaint refers to the specific device being investigated for this complaint record.Dhr review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.The customer did not submit images for the reported event.Based on the investigation and risk management file review, the most likely root causes determined from the investigation are parafunctional habits/patient factors, or customer error.Therefore, based on the available information, a device malfunction did occur.The reported event was confirmed with all the available information.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMP,TSV,MCOL MG,6.0MM,8MM,MTX
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key23000602
Report Number0002023141-2025-02457
Device Sequence Number8869038
Product Code DZE
UDI-Device Identifier00889024019713
UDI-Public(01)00889024019713(17)241020(10)1233088
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K111889
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date10/20/2024
Device Catalogue NumberTSVM6B8
Device Lot Number1233088
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/13/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date09/09/2025
Supplement Report FDA Received Date10/17/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/22/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
Patient SexFemale