• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES LLC CARPENTIER-EDWARDS PERIMOUNT MAGNA EASE PERICARDIAL AORTIC BIOPROSTHESIS; REPLACEMENT HEART VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES LLC CARPENTIER-EDWARDS PERIMOUNT MAGNA EASE PERICARDIAL AORTIC BIOPROSTHESIS; REPLACEMENT HEART VALVE Back to Search Results
Model Number 3300TFX
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Chest Pain (1776); Insufficient Information (4580)
Date of Event 07/28/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H11: additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event or Problem Description
It was learned through implant patient registry that a patient with a 27mm 3300tfx aortic valve was explanted after an implant duration of 6 years, 4 months due to unknown reasons.The explanted valve was replaced with a 27mm 11500a inspiris resilia valve.Edwards lifesciences maintains an implant patient registry.This registry is a patient tracking mechanism for serialized edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true post-market surveillance registry.Through the registry, edwards is notified when these devices are implanted.In addition, patient and/or device status may be reported to the registry via the implantation data cards.The information is received from various sources (e.G.Surgeon, hospital, and patient family members) and is not received in the form of a conventional 'customer complaint'.The information reported may or may not be related to the edwards device.
 
Event or Problem Description
It was learned through implant patient registry and investigation that a patient with a 27mm 3300tfx aortic valve was explanted after an implant duration of 6 years, 4 months due to pannus.Patient presented with chest pain.The explanted valve was replaced with a 27mm 11500a inspiris resilia valve.Patient was sent to recovery and discharged on pod#8.This is a 40-year-old male patient: edwards lifesciences maintains an implant patient registry.This registry is a patient tracking mechanism for serialized edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true post-market surveillance registry.Through the registry, edwards is notified when these devices are implanted.In addition, patient and/or device status may be reported to the registry via the implantation data cards.The information is received from various sources (e.G.Surgeon, hospital, and patient family members) and is not received in the form of a conventional 'customer complaint'.The information reported may or may not be related to the edwards device.
 
Additional Manufacturer Narrative
H11: additional manufacturer narrative: updated sections: b7, g3, g6, h2, h6 (type of investigation) corrected sections: b5, h6 (component code, clinical code, device code, investigation findings and investigation conclusions).Pannus is a non-immune inflammatory reaction of the body to the implanted prosthesis, characterized by proliferation of fibroelastic tissue and collagen, with a starting point in the suture area and subacute or chronic centripetal evolution.Pannus can have both beneficial and harmful effects depending on the amount of growth and location.A small amount of host tissue growth over the suture line is expected and is needed to form a non-thrombogenic surface and complete the healing process after device implantation.In contrast, an excessive amount of pannus growth can cause nonstructural dysfunction that may require intervention.Host tissue growth can extend onto the cusp surfaces leading to thickening of the cusps, cusp retraction or curling, leaflet immobility, and/or abnormal coaptation, potentially resulting in valvular regurgitation, elevated gradients, and/or stenosis.Pannus/host tissue overgrowth is most likely due to patient factors and is unlikely to be related to a manufacturing non-conformance.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.The most likely root cause is patient factors.A device history record (dhr) review is not performed because there is no allegation of a malfunction which could be related to a manufacturing non-conformance and/or one was not suspected or confirmed through investigation; no labeling non-conformance/deficiency; no device-related infection; and no evidence of a product failure with regard to design, reliability, or use error.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT MAGNA EASE PERICARDIAL AORTIC BIOPROSTHESIS
Common Device Name
REPLACEMENT HEART VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES LLC
one edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES LLC
one edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
mle fl2 - office m2013
irvine, CA 92614
9492506615
MDR Report Key23000807
Report Number2015691-2025-07392
Device Sequence Number18391836
Product Code DYE
UDI-Device Identifier00690103176162
UDI-Public(01)00690103176162(17)221001
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P860057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/01/2022
Device Model Number3300TFX
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/14/2025
Supplement Date Received by Manufacturer10/15/2025
Initial Report FDA Received Date09/09/2025
Supplement Report FDA Received Date10/20/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/02/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention; Life Threatening;
Patient Age40 YR
Patient SexMale
-
-