MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD POLYSORB; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC AC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD POLYSORB; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC AC Back to Search Results
Model Number UL-202
Medical Device Problem Codes Break (1069); Product Quality Problem (1506)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/15/2025
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter during a laparoscopic general surgery, when opening the package it was discolored and when pulled out, it broke.It occurred on ten sutures.The same product in another box was used to resolve the issue.There was no patient injury.
 
Additional Manufacturer Narrative
D10 concomitant medical products: ul-202, ul-202 polysorb* 5-0 vio 75cm cv23 x36 (lot#:d4j2565y), ul-202, ul-202 polysorb* 5-0 vio 75cm cv23 x36 (lot#:d4j2565y), ul-202, ul-202 polysorb* 5-0 vio 75cm cv23 x36 (lot#:d4j2565y), ul-202, ul-202 polysorb* 5-0 vio 75cm cv23 x36 (lot#:d4j2565y), ul-202, ul-202 polysorb* 5-0 vio 75cm cv23 x36 (lot#:d4j2565y), ul-202, ul-202 polysorb* 5-0 vio 75cm cv23 x36 (lot#:d4j2565y), ul-202, ul-202 polysorb* 5-0 vio 75cm cv23 x36 (lot#:d4j2565y), ul-202, ul-202 polysorb* 5-0 vio 75cm cv23 x36 (lot#:d4j2565y), ul-202, ul-202 polysorb* 5-0 vio 75cm cv23 x36 (lot#:d4j2565y) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.Ten sealed representative samples were returned for evaluation.The evaluation found no potentially contributing factors.It was reported that during a laparoscopic general surgery, when opening the package, the suture was found discolored, and it broke when it was pulled out.The reported issues could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYSORB
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC AC
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR   0101
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key23001575
Report Number9612501-2025-02384
Device Sequence Number21104704
Product Code GAM
UDI-Device Identifier10884521049444
UDI-Public(01)10884521049444(17)290831(10)D4J2565Y(20)01
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K963253
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberUL-202
Device Catalogue NumberUL-202
Device Lot NumberD4J2565Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/15/2025
Supplement Date Received by Manufacturer11/06/2025
Initial Report FDA Received Date09/09/2025
Supplement Report FDA Received Date11/10/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/05/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Patient SexUnknown
-
-