| Model Number |
UL-202 |
| Medical Device Problem Codes |
Break (1069); Product Quality Problem (1506)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/15/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter during a laparoscopic general surgery, when opening the package it was discolored and when pulled out, it broke.It occurred on ten sutures.The same product in another box was used to resolve the issue.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant products: ul-202, ul-202 polysorb5-0 vio 75cm cv23 x36 (lot 5-0 vio 75cm cv23 x36 (lot#:d4j2565y), ul-202, ul-202 polysorb* 5-0 vio 75cm cv23 x36 (lot#:d4j2565y), ul-202, ul-202 polysorb* 5-0 vio 75cm cv23 x36 (lot#:d4j2565y), ul-202, ul-202 polysorb* 5-0 vio 75cm cv23 x36 (lot#:d4j2565y), ul-202, ul-202 polysorb 5-0 vio 75cm cv23 x36 (lot#:d4j2565y), ul-202, ul-202 polysorb 5-0 vio 75cm cv23 x36 (lot#:d4j2565y), ul-202, ul-202 polysorb 5-0 vio 75cm cv23 x36 (lot#:d4j2565y) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.Ten sealed representative samples were returned for evaluation.The evaluation found no potentially contributing factors.It was reported that during a laparoscopic general surgery, when opening the package, the suture was found discolored, and it broke when it was pulled out.The reported issues could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|