MAUDE Adverse Event Report: KEYSTONE DENTAL, INC. PRMA+; PRIMA PLUS IMPLANT 5.0X8.0 MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KEYSTONE DENTAL, INC. PRMA+; PRIMA PLUS IMPLANT 5.0X8.0 MM Back to Search Results
Model Number 15740K
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 08/25/2025
Type of Reportable Event Serious Injury
Event or Problem Description
Failed to osseointegrate.
 
Additional Manufacturer Narrative
Keystone dental inc.Provides labeling with all implants.Failure to osseointegrate and loss of osseointegration (implant mobility) is identified as an adverse reaction and identified in the list of warnings in all endosseous dental implant labeling.The specific cause for a particular complaint is often not readily identified as there are various factors that contribute to the risk of implant failure.These include patient factors such as prior oral infection, poor bone quality or quantity, systemic conditions such as diabetes and uncontrolled hypertension.Technique factors such as overheating of the implant site may cause necrosis of the bone, preventing growth (pelayo et al., 2008; turkyilmaz & al., 2008).Patient habits such as tobacco use (heitz-mayfield & huynh-ba, 2009), alcohol or drug abuse, poor oral hygiene, and bruxism (salvi & braegger, 2009) may also lead to implant failure.In addition, improper surgical technique can lead to implant failure and/or loss of supporting bone (salvi & braegger, 2009).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRMA+
Common Device Name
PRIMA PLUS IMPLANT 5.0X8.0 MM
Manufacturer (Section D)
KEYSTONE DENTAL, INC.
13645 alton pkwy ste a
irvine CA 92618
Manufacturer (Section G)
KEYSTONE DENTAL, INC.
13645 alton pkwy ste a
irvine CA 92618
Manufacturer Contact
kathleen labrie
13645 alton pkwy ste a
irvine, CA 92618
9784308257
MDR Report Key23003053
Report Number3005990499-2025-71788
Device Sequence Number18521392
Product Code DZE
UDI-Device IdentifierD76815740K0
UDI-Public+D76815740K0/$$3251015WO014895T
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Dentist
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 09/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/15/2025
Device Model Number15740K
Device Catalogue Number15740K
Device Lot NumberWO-014895
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/26/2025
Initial Report FDA Received Date09/09/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/18/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age30 YR
Patient SexMale
-
-