Philips has investigated this complaint.According to additional information collected, the system was not in clinical use when the issue occurred.The philips field service engineer (fse) visited system onsite and confirmed the x-ray tube was failed.Upon troubleshooting, the fse identified that the tube was leaking the oil.To resolve the issue, the fse replaced the tube and returned the defective part for further analysis.The supplier¿s part analysis could not determine the cause of the tube failure, but the insert shows normal signs of wear and tear.After the replacement of tube, the system was returned to use in good working order.At the time this complaint was received, philips did not have enough information to exclude the potential for death or serious injury on recurrence, and as such the complaint was reported.Since that time, new information has been received which has led to the determination that this is scenario is not likely to cause or contribute to death or serious injury if it were to recur.As per the instructions for use, the user of the product must institute a user routine checks program.The user of the product shall make sure that all checks and actions have been satisfactorily completed before using the product for its intended purpose.It is instructed to not use the product for any application until the user is sure that the user routine checks have been satisfactorily completed, therefore, as the device was outside of use at the time the issue was identified, the user would identify this issue prior to use.Based on re-evaluation, this case is not reportable.The codes were updated based on the investigation outcome.
|