Model Number AZURION 7 M20
Medical Device Problem Code Radiation Output Failure (4027)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/12/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to philips that the system's x-ray tube failed, which can impact imaging.No harm to the patient or user was reported.Philips has started an investigation of this complaint.
 
Additional Manufacturer Narrative
Philips has investigated this complaint.According to additional information collected, the system was not in clinical use when the issue occurred.The philips field service engineer (fse) visited system onsite and confirmed the x-ray tube was failed.Upon troubleshooting, the fse identified that the tube was leaking the oil.To resolve the issue, the fse replaced the tube and returned the defective part for further analysis.The supplier¿s part analysis could not determine the cause of the tube failure, but the insert shows normal signs of wear and tear.After the replacement of tube, the system was returned to use in good working order.At the time this complaint was received, philips did not have enough information to exclude the potential for death or serious injury on recurrence, and as such the complaint was reported.Since that time, new information has been received which has led to the determination that this is scenario is not likely to cause or contribute to death or serious injury if it were to recur.As per the instructions for use, the user of the product must institute a user routine checks program.The user of the product shall make sure that all checks and actions have been satisfactorily completed before using the product for its intended purpose.It is instructed to not use the product for any application until the user is sure that the user routine checks have been satisfactorily completed, therefore, as the device was outside of use at the time the issue was identified, the user would identify this issue prior to use.Based on re-evaluation, this case is not reportable.The codes were updated based on the investigation outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AZURION
Common Device Name
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL   5684 PC
Manufacturer Contact
marjan vos
veenpluis 6
best 5684 -PC
NL   5684 PC
MDR Report Key23003148
Report Number3003768277-2025-009468
Device Sequence Number15702617
Product Code OWB
UDI-Device Identifier00884838101883
UDI-Public(01)00884838101883(10)3.0
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K200917
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 02/04/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAZURION 7 M20
Device Catalogue Number722224
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/12/2025
Supplement Date Received by Manufacturer02/02/2026
Initial Report FDA Received Date09/09/2025
Supplement Report FDA Received Date02/04/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/17/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1