| Model Number |
CCW0T5 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device was not returned for analysis.A device history record review and a non-conformance review of the reported lot number was conducted.No deviations were identified and all corresponding production release specifications defined in the device master record were met.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|
Event or Problem Description
|
|
A non-healthcare professional reported during intraocular lens (iol) implant procedure, when lens was loaded and placed in the eye, a crack was noticed from one side to other.The lens was explanted during initial procedure.A second lens was implanted, and it also had a crack in the exact same place all the way across the lens.Yet, this one was left implanted.Additional information was requested.
|
| |
|
Event or Problem Description
|
|
Additional information was received stating that the current lens was only opened and examined under microscope and no damage was found.There is no issue with the lens.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information has been provided in a.1., b.5.And h.11.Based on information received following submission of the initial report, there is no issue with the lens and does not meet criteria for reporting as a malfunction.No further reports required.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|