Catalog Number A4E06
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/11/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was received for investigation, and the reported problem was confirmed.The balloon was observed to be ruptured.While the event was verified, the exact root cause could not be determined.Balloon rupture is an inherent risk associated with this type of device, and given the nature of the product and the procedures in which it is used, this event is considered a known complication.Procedural factors or anatomical features, such as calcification or plaque, may have contributed to the reported issue.As stated in the instructions for use (ifu), complications may occur during the use of embolectomy catheters.These may include, but are not limited to: intimal disruption, vessel wall perforation, balloon rupture with fragmentation, tip separation, local or systemic infection, arterial thrombosis, local hematomas, air embolus, rupture of an aneurysm, arterial spasm, distal embolus of blood clots or arteriosclerotic plaque, arterial dissection, and hemorrhage.Arterial rupture or balloon failure due to sharp calcified plaque may also occur.The possibility of balloon rupture must be taken into account when evaluating the risks of catheterization procedures.A lot history review was conducted, and no issues were identified in the manufacturing or packaging processes that could be related to this event.All quality control testing met specifications.
 
Event or Problem Description
It was reported that the syntel over-the-wire catheter balloon ruptured during patient use.The device had been checked prior to use.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNTEL SILICONE OVER-THE-WIRE EMBOLECTOMY CATHETER
Common Device Name
EMBOLECTOMY CATHETER
Manufacturer (Section D)
LEMAITRE VASCULAR, INC.
63 second ave
burlington MA 01803
Manufacturer (Section G)
LEMAITRE VASCULAR, INC
63 second ave
burlington MA 01803
Manufacturer Contact
peter song
63 second ave
burlington, MA 01803
7814251683
MDR Report Key23003497
Report Number1220948-2025-00153
Device Sequence Number15857889
Product Code DXE
UDI-Device Identifier00840663109913
UDI-Public(01)00840663109913(11)240820(17)290728(10)PYT00001308(91)A4E06
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K202049
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 09/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberA4E06
Device Lot NumberPYT00001308
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/28/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/11/2025
Initial Report FDA Received Date09/09/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/20/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-