| Model Number |
LOW PROFILE SCREW, 1.0X10MM, CORTICAL,TI |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/08/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2025, it was reported by a sales representative via (b)(4) that an ar-18710-10 low profile screw broke while being implanted.This was discovered during the case with no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.Complaint allegation is not confirmed.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause of the reported failure is user error in the device, which can be attributed to prying/leveraging, misalignment, and/or excessive force used during implant insertion.
|
| |
|