STRYKER GMBH WIRE WITH OLIVE, DIAMOND POINT HOFFMANN LRF Ø1.8 X 450MM; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT Back to Search Results
Catalog Number 49338030
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/11/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Analysis results are not available at the time of this report.A follow-up report will be sent when the analysis is complete.
 
Event or Problem Description
It was reported by the customer that there were four wire breakages on a frame.The wires broke at the interface with the wire bolt.The primary surgery was completed successfully.The patient was weight bearing and there were no falls or trips to cause the reported event.During the revision surgery it was reported that another wire in the proximal ring at the olive broke.
 
Event or Problem Description
It was reported by the customer that there were four wire breakages on a frame.The wires broke at the interface with the wire bolt.The primary surgery was completed successfully.The patient was weight bearing and there were no falls or trips to cause the reported event.During the revision surgery it was reported that another wire in the proximal ring at the olive broke.
 
Additional Manufacturer Narrative
Correction to d9 (product available to stryker) and h6(device code).The complaint couldn't be confirmed, since the returned image for evaluation don't matches the alleged failure.The investigation revealed the following: following our inspection of the x-ray images, we can confirm that one wire is no longer connected to the wire bolt.However, it is not possible to determine from the images whether the wire is broken or simply disconnected.No other abnormalities are visible.For a comprehensive evaluation, we kindly request that the material be returned to us.A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.Indications of material, manufacturing, or design related problems were unable to be identified as the lot number was not communicated.More detailed information about the complaint event must be available in order to determine the root cause of the complaint event.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WIRE WITH OLIVE, DIAMOND POINT HOFFMANN LRF Ø1.8 X 450MM
Common Device Name
APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ   2545
Manufacturer Contact
dannielle stap
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key23005063
Report Number0008031020-2025-01468
Device Sequence Number9847991
Product Code KTT
UDI-Device Identifier07613252611011
UDI-Public07613252611011
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K233741
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number49338030
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/11/2025
Supplement Date Received by Manufacturer10/15/2025
Initial Report FDA Received Date09/09/2025
Supplement Report FDA Received Date11/12/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age30 YR
Patient SexUnknown
Patient Weight72 KG
-
-