| Catalog Number |
511.791 |
| Medical Device Problem Codes |
Device-Device Incompatibility (2919); Compatibility Problem (2960)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/05/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was noted in the service order that the quick coupling device had a spit pusher that no longer held the spits.During service and evaluation, it was determined that the device had damages/breaks k-wire.It was further determined that the device failed pretest for check tool coupling, check self-holding force in smallest range and check function in running mode.This event did not occur during surgery.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H11: additional narrative: device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the device damages/breaks k-wire, identified during service and evaluation was confirmed.The assignable root cause was determined to be traced to maintenance, which is improper maintenance.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H11: additional narrative: correction.B5 and h6 updated.
|
| |
|
Event or Problem Description
|
|
It was noted in the service order that the quick coupling device had a spit pusher that no longer held the spits.During service and evaluation, it was determined that the device had damages/breaks k-wire and will not hold the k-wire.It was further determined that the device failed pretest for check tool coupling, check self-holding force in smallest range and check function in running mode.This event did not occur during surgery.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.
|
| |
|