RESPIRONICS, INC. SIMPLYGO MINI; GENERATOR, OXYGEN, PORTABLE Back to Search Results
Model Number 1113603
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/26/2024
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging that the simplygo mini device would not charge.The device was not in use on a patient at the time of the occurrence.The device was returned to the philips service center for evaluation, and the customer¿s complaint was confirmed.During the evaluation, it was noted that the device had a burnt main pca.In conclusion, the device's pca was replaced to address the issue.Additionally, during the service of the device, the inlet filters were replaced, the sieves were replaced, the air side assembly was noisy and replaced, and the o2 side assembly was replaced.These were unrelated to the reportable malfunction/issue.The device passed all final testing.At this time, no further investigation can be performed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIMPLYGO MINI
Common Device Name
GENERATOR, OXYGEN, PORTABLE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key23007302
Report Number2518422-2025-109847
Device Sequence Number13383097
Product Code CAW
UDI-Device Identifier00606959032415
UDI-Public(01)00606959032415
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K111885
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number1113603
Device Catalogue Number1113603
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/10/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/10/2024
Initial Report FDA Received Date09/09/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/11/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-