Model Number FXS4012C
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Codes Failure of Implant (1924); Inadequate Osseointegration (2646)
Date of Event 05/06/2024
Type of Reportable Event Serious Injury
Event or Problem Description
The doctor reported the implant was removed due to osseointegration failure within 1 year, approximately 10 months after implant placement.The bone quality was type iii.The oral hygiene around implant site was poor.Bone resorption and peri-implantitis were observed during the removal surgery.Bone augmentation material was used in implant placement surgery.The patient will need another surgical intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERLINE II
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
DENTIUM USA
6761 katella ave
cypress CA 90630
Manufacturer (Section G)
DENTIUM USA
6761 katella ave
cypress CA 90630
Manufacturer Contact
eun kyung son
6761 katella ave
cypress, CA 90630
7142260229
MDR Report Key23007864
Report Number3005503242-2025-00112
Device Sequence Number9702152
Product Code DZE
UDI-Device IdentifierD755FXS4012C0
UDI-PublicD755FXS4012C0
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K160965
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 09/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberFXS4012C
Device Catalogue NumberFXS4012C
Device Lot NumberH25LB905S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/19/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/19/2025
Initial Report FDA Received Date09/09/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/29/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
Patient SexFemale
-
-