MAQUET CARDIOVASCULAR LLC HEARTSTRING III SYSTEM 4.3MM; CLAMP, VASCULAR Back to Search Results
Model Number HSK-3043
Medical Device Problem Code Fitting Problem (2183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/20/2025
Type of Reportable Event Malfunction
Event or Problem Description
The hospital reported that they had a hst iii system (4.3 mm) misfire.When pulling the plunger from the loading device, the seal did not properly load in the delivery tube.Opened a new kit to complete the procedure.Only delay was length of time to open a new kit.No patient injury.
 
Additional Manufacturer Narrative
Tw (b)(4).The device has not yet been returned to maquet cardiac surgery for evaluation.We are following up with the customer for the return of the device.A supplemental report will be submitted if the device is received.
 
Additional Manufacturer Narrative
Trackwise - (b)(4).Updated field: b4, d9, g3, g6, h2, h3, h6 (type of investigation, investigation findings, component codes, investigation conclusions), h11.Corrected field: d1.The lot # 3000479466 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTSTRING III SYSTEM 4.3MM
Common Device Name
CLAMP, VASCULAR
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 07470
MDR Report Key23008808
Report Number2242352-2025-0001048
Device Sequence Number21258210
Product Code DXC
UDI-Device Identifier00607567700321
UDI-Public(01)00607567700321(10)3000479466(17)260520
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K130382
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 10/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/20/2026
Device Model NumberHSK-3043
Device Catalogue NumberHSK-3043
Device Lot Number3000479466
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/21/2025
Supplement Date Received by Manufacturer10/30/2025
Initial Report FDA Received Date09/09/2025
Supplement Report FDA Received Date10/30/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/20/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN.
Patient Age67 YR
Patient SexMale
Patient Weight75 KG
-
-