| Catalog Number |
368380 |
| Medical Device Problem Code |
Device Ingredient or Reagent Problem (2910)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/15/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® trace element serum blood collection tubes, improper separation was seen in one tube resulting in sample recollection.No adverse impact was reported regarding the patient or user.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® trace element serum blood collection tubes, improper separation was seen in one tube resulting in sample recollection.No adverse impact was reported regarding the patient or user.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary bd received three photos for investigation, which showed the presence of poor separation.Additionally, 100 retained samples were visually inspected, with no issues being identified.Complaints for sample quality are under statistical control for the month of august 2025.At this time, further testing is not indicated.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has been confirmed for the indicated failure mode: poor barrier separation.No definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.Bd quality will continue to monitor sample quality complaints.
|
| |
|