Catalog Number 320-10-00
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Pain (1994)
Date of Event 08/20/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D10: (b)(6), 300-01-15 - equinoxe, humeral stem primary, press fit 15mm, (b)(6), 320-01-42 - equinoxe reverse 42mm glenosphere, (b)(6), 320-10-00 - equinoxe reverse tray adapter plate tray +0, (b)(6), 320-15-01 - eq rev glenoid plate, (b)(6), 320-15-05 - eq rev locking screw, (b)(6), 320-20-00 - eq reverse torque defining screw kit, (b)(6), 320-20-22 - eq rev compress screw lck cap kit, 4.5 x 22mm, (b)(6), 320-20-22 - eq rev compress screw lck cap kit, 4.5 x 22mm, (b)(6), 320-20-26 - eq rev compress screw lck cap kit, 4.5 x 26mm, (b)(6), 320-20-30 - eq rev compress screw lck cap kit, 4.5 x 30mm, (b)(6), 320-42-10 - equinoxe reverse 42mm humeral const liner +0, (b)(6), 321-20-00 - equinoxe reverse shoulder drill kit (b)(6),300-01-15 - equinoxe, humeral stem primary, press fit.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Event or Problem Description
It was reported that the patient underwent an initial right reverse total shoulder arthroplasty.Subsequently, the patient experienced pain and atraumatic disassociation of the humeral liner and humeral tray.As a result, the patient underwent a shoulder revision approximately 6 years and 6 months after initial implantation.Patient was last known to be in stable condition following the event.No further patient impact reported.No further information available.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: h6.Mdr section codes updated/corrected: b, c.The revision reported was likely the result of disassembly between the humeral liner and humeral tray.However, this cannot be confirmed and potential contributions from patient conditions, an unreported traumatic event, incomplete seating of the humeral liner during implantation, forceful contact between the liner and glenoid bone (scapular notching), or a combination of the above to the reported disassembly of the humeral liner cannot be determined as the devices were not returned for evaluation and images were not provided.Should additional relevant information be obtained, a follow-up mdr will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE REVERSE TRAY ADAPTER PLATE TRAY +0
Common Device Name
PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC
2320 nw 66th ct.
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
MDR Report Key23009686
Report Number1038671-2025-02903
Device Sequence Number10559550
Product Code KWT
UDI-Device Identifier10885862086419
UDI-Public10885862086419
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K063569
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number320-10-00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/20/2025
Supplement Date Received by Manufacturer10/10/2025
Initial Report FDA Received Date09/09/2025
Supplement Report FDA Received Date10/15/2025
Date Device Manufactured12/27/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age72 YR
Patient SexMale
-
-