| Model Number |
420205-16 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/28/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The instrument involved in this complaint has not been received and/or tested by failure analysis as of the date of this report.If the product is received and evaluated, a supplemental mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a broken cable with a da vinci product.The customer reported event has no report of patient injury.
|
| |
|
Event or Problem Description
|
|
Refer to h11 for follow-up information.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) received the da vinci product to perform failure analysis.The fenestrated bipolar forceps instrument was analyzed and found to have a derailed pitch cable at the proximal end.Common causes of the failure mode derailed instrument pitch cables are attributed to a component failure.Derailed cables are attributed to a loss of cable tension.A cracked clamping pulley and/or input shaft can cause the cable to become dislodged at the proximal end.When the clamping pulley and/or input shaft is cracked, the cable can dislodge as the input could "slip" with respect to its internal shaft causing the loss of cable tension.A cracked clamping pulley and/or input shaft can be caused by improper cleaning or sterilization techniques used during reprocessing.A dislodged pitch cable at the proximal end can then result in the cable becoming derailed or loose at the distal end, which may lead to non-intuitive motion.
|
| |
|