W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® CONFORMABLE AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
|
Back to Search Results |
|
| Catalog Number |
CXT231412 |
| Medical Device Problem Codes |
Material Deformation (2976); Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/26/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 cfr part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Event or Problem Description
|
|
The following information was reported to gore: on (b)(6) 2025, the patient underwent endovascular treatment for an abdominal aortic aneurysm using the gore® excluder® conformable aaa endoprosthesis, and the gore® excluder® aaa endoprosthesis.Reportedly, the patient¿s common iliac arteries showed aneurysmal changes.During deployment of the trunk-ipsilateral leg endoprosthesis, the long marker was unintentionally placed at the level of the renal arteries, requiring repositioning using the constraining system.During this process, the contralateral gate became caught within the aneurysm sac and was compressed.The delivery catheter was operated for adjustment, and the proximal portion was redeployed, but full expansion was not achieved.Cannulation of the contralateral gate was then performed while the delivery catheter was held with forceps.After deployment of all stent graft components, touch-up of the right limb was performed using a non-gore balloon in the right common iliac artery, during which a rupture was observed.To repair the injury, two additional stent grafts were implanted.Physician's comment: excessive pressure was applied during the touch-up.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: a review of the manufacturing records for the device verified that the lot met all pre-release specifications.
|
| |
|