BIOMET UK LTD. CER BIOLOXD OPTION HD 28MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER Back to Search Results
Catalog Number 650-1055
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Joint Laxity (4526); Medical device site infection (4845); Periprosthetic fracture (4902)
Date of Event 03/20/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that patient underwent to hip revision surgery approximately two months post-implantation due to infection and instability.During the revision, a fracture of the greater trochanter occurred due to the amount of lysis from prior metal on metal hip replacement.
 
Additional Manufacturer Narrative
(b)(4).D10: cer option type 1 tpr sleve +6; item number# 650-1068; lot number# 3207905.Act artic hd arcom xl 28x38mm; item number# xl-200144; lot number# 66747541.Unknown stem; item number# unknown; lot number# unknown.Unknown shell; item number# unknown; lot number# unknown.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated: b4, b5, d2a, g3, g6, h2, h3, h6, h10, h11.The products involved in the reported event were not returned for investigation; however, pictures were provided and reviewed.The explanted products showed the presence of bio-debris; however, this observation cannot be used to confirm the reported event.As the devices were not returned, no further analysis can be performed.A review of the device manufacturing records confirmed no abnormalities or deviations.Review of the complaint histories identified additional similar complaints for the reported items and no additional similar complaints for the reported part and lot combinations.Medical records were provided and reviewed by a health care professional.The review identified the components were revised due to infection, instability, and prosthetic failure.We did have complications with a greater trochanteric fracture from the amount of lysis from his prior metal-on-metal hip replacement.A definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CER BIOLOXD OPTION HD 28MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key23013649
Report Number3002806535-2025-00448
Device Sequence Number9769812
Product Code LZO
UDI-Device Identifier00887868271373
UDI-Public(01)00887868271373(17)341014(10)3216983
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K200959
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue Number650-1055
Device Lot Number3216983
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/18/2025
Supplement Date Received by Manufacturer01/09/2026
Initial Report FDA Received Date09/10/2025
Supplement Report FDA Received Date01/09/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/14/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexMale
Patient Weight95 KG
Patient EthnicityNon Hispanic
-
-